For industrial ozone applications not serviced by the UV-100, Eco Sensors has partnered with 2B Technologies to offer the Model 106-L, 106-M, 106-MH and 106-H Ozone Monitors, where L, M, MH and H refer to Low (0-100 ppm), Medium (0-1,000 ppm), Medium-High (0-10,000 ppm) and High (0-20 wt%) ozone concentrations, respectively. These instruments have different optical path lengths in order to accommodate an extremely wide range of ozone concentrations, spanning more than eight orders of magnitude. Additionally, the Model 106-H was designed with a flow through path that can be pressurized to up to 50 psi for in-line measurements with ozone generators. All three models are available in either a standard enclosure or NEMA enclosure. Please call us for pricing and availability.
Specifications for Different Models
Model | Range | Resolution | Precision & Accuracy |
---|---|---|---|
106-L
|
0-100 ppm
|
0.0001 ppm (0.1 ppb)
|
Higher of 0.0015 ppm (1.5 ppb) or 2% of reading
|
106-M
|
0-1,000 ppm
|
0.01 ppm
|
Higher of 0.01 ppm or 2% of reading
|
106-MH
|
0-10,000 ppm (0-1 vol%)
|
0.1 ppm
|
Higher of 0.05 ppm or 2% of reading
|
106-H
|
0-20 wt% O2 or Air
0-14 vol% |
0.01 wt%
0.01 vol% |
Higher of 0.01 wt% or 2% of reading
|
General Specifications
Measurement Principle | UV Absorption at 254 nm |
Measurement Interval | 2 s |
Data Averaging Options | 10 s, 1 min, 5 min, 1 hr |
Flow Rate | ~1 Liter/min; Up to 20 L/min for Model 106-H |
Data Storage | 32,736 lines (10 s avg. = 3.7 days; 5 min avg = 113 days) |
Choice of Units | ppb, ppm, pphm, µg m-3, mg m-3 (for Models 106-L and 106-M) ppm, mg m-3 (for Model 106-MH) vol%, wt% air, wt% O2, g/m3, g/Nm3 (for Model 106-H) |
Data Outputs | USB, RS232, 0-2.5 V Analog, 4-20 mA, LCD Display |
Power Requirements | 12 V, 500 mA, 6 watt |
Size | 3.75 x 8.5 x 8.5 inches (9.5 x 21.6 x 21.6 cm) |
Weight | 4.5 lb (2.0 kg) |
Features
- Measurement based on UV absorption
- Low power consumption; can be battery operated
- Internal data logger with real time clock
- On-board microprocessor with interactive menus includes data averaging options of 10 s, 1 m, 5 m and 1 hr
- USB and RS-232 output of time/date, O3 concentration, internal temperature and pressure
- Analog output (0-2.5 V and 4-20 mA) of ozone concentration in user selected units
- Two 2-level relays for control purposes (e.g., control of ozone source or turn warning light on and off)
- Long-life pump (15,000 hr) for Models 106-L, 106-M and 106-MH
To download the product manuals for the UV-106 product line, click on the links below.